Pulse oximeter method as an indicator of blood perfusion in neonate patients / Naszariah Mohd Noor
Blood gas measurements normally provide critical information on oxygenation, ventilation and acid-base status. In critically ill neonates, where peripheral perfusion may drop below the minimums required for tissue oxygenation and cellular respiration, the risk associated with rapid changes in oxy...
Saved in:
Main Author: | |
---|---|
Format: | Thesis |
Language: | English |
Published: |
2013
|
Subjects: | |
Online Access: | https://ir.uitm.edu.my/id/eprint/12024/1/TM_NASZARIAH%20MOHD%20NOOR%20EE%2013_5%201.pdf |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
id |
my-uitm-ir.12024 |
---|---|
record_format |
uketd_dc |
spelling |
my-uitm-ir.120242016-05-14T07:43:43Z Pulse oximeter method as an indicator of blood perfusion in neonate patients / Naszariah Mohd Noor 2013 Mohd Noor, Naszariah TK Electrical engineering. Electronics. Nuclear engineering Blood gas measurements normally provide critical information on oxygenation, ventilation and acid-base status. In critically ill neonates, where peripheral perfusion may drop below the minimums required for tissue oxygenation and cellular respiration, the risk associated with rapid changes in oxygen saturation is high. Thus, effective means of monitoring blood oxygen saturation is essential. Using pulse oximetry is appealing because it is non-invasive and correlates well with arterial gases. Optimal pulse oximetry monitoring accuracy is depending on the selection of a monitoring site characterized by good perfusion with oxygen saturation. The aim of this research is to investigate the use of pulse oximetry as a non-invasive indicator of blood perfusion of preterm newborns. One hundred and fifty (89 male and 61 female) preterm newborns, regardless of ethnic group and delivery norm, admitted in Special Care Nursery University Malaya Medical Centre were recruited for this observational study from 2006 to 2008. The inclusion criteria were neonates with a diagnosis of respiratory distress syndrome who required ventilatory support, gestational age between 28 weeks to 37 weeks, and weighing less than 2500g. Healthy preterm neonates weighing more than 2500g were excluded. Mathematical modeling was used to model the correlation between the oxygen saturation and blood perfusion. With the analysis made, perfusion index correlates positively with oxygen saturation for both palm and foot sensors (r = 0.91, r = 0.92) respectively with an accuracy of 96%. Hence, pulse oximetry technique is a useful tool to indicate the blood perfusion in neonate patients. 2013 Thesis https://ir.uitm.edu.my/id/eprint/12024/ https://ir.uitm.edu.my/id/eprint/12024/1/TM_NASZARIAH%20MOHD%20NOOR%20EE%2013_5%201.pdf text en public mphil masters Universiti Teknologi MARA Perpustakaan Tun Abdul Razak |
institution |
Universiti Teknologi MARA |
collection |
UiTM Institutional Repository |
language |
English |
topic |
TK Electrical engineering Electronics Nuclear engineering |
spellingShingle |
TK Electrical engineering Electronics Nuclear engineering Mohd Noor, Naszariah Pulse oximeter method as an indicator of blood perfusion in neonate patients / Naszariah Mohd Noor |
description |
Blood gas measurements normally provide critical information on oxygenation,
ventilation and acid-base status. In critically ill neonates, where peripheral perfusion
may drop below the minimums required for tissue oxygenation and cellular
respiration, the risk associated with rapid changes in oxygen saturation is high. Thus,
effective means of monitoring blood oxygen saturation is essential. Using pulse
oximetry is appealing because it is non-invasive and correlates well with arterial
gases. Optimal pulse oximetry monitoring accuracy is depending on the selection of a
monitoring site characterized by good perfusion with oxygen saturation. The aim of
this research is to investigate the use of pulse oximetry as a non-invasive indicator of
blood perfusion of preterm newborns. One hundred and fifty (89 male and 61 female)
preterm newborns, regardless of ethnic group and delivery norm, admitted in Special
Care Nursery University Malaya Medical Centre were recruited for this observational
study from 2006 to 2008. The inclusion criteria were neonates with a diagnosis of
respiratory distress syndrome who required ventilatory support, gestational age
between 28 weeks to 37 weeks, and weighing less than 2500g. Healthy preterm
neonates weighing more than 2500g were excluded. Mathematical modeling was
used to model the correlation between the oxygen saturation and blood perfusion.
With the analysis made, perfusion index correlates positively with oxygen saturation
for both palm and foot sensors (r = 0.91, r = 0.92) respectively with an accuracy of
96%. Hence, pulse oximetry technique is a useful tool to indicate the blood perfusion
in neonate patients. |
format |
Thesis |
qualification_name |
Master of Philosophy (M.Phil.) |
qualification_level |
Master's degree |
author |
Mohd Noor, Naszariah |
author_facet |
Mohd Noor, Naszariah |
author_sort |
Mohd Noor, Naszariah |
title |
Pulse oximeter method as an indicator of blood perfusion in neonate patients / Naszariah Mohd Noor |
title_short |
Pulse oximeter method as an indicator of blood perfusion in neonate patients / Naszariah Mohd Noor |
title_full |
Pulse oximeter method as an indicator of blood perfusion in neonate patients / Naszariah Mohd Noor |
title_fullStr |
Pulse oximeter method as an indicator of blood perfusion in neonate patients / Naszariah Mohd Noor |
title_full_unstemmed |
Pulse oximeter method as an indicator of blood perfusion in neonate patients / Naszariah Mohd Noor |
title_sort |
pulse oximeter method as an indicator of blood perfusion in neonate patients / naszariah mohd noor |
granting_institution |
Universiti Teknologi MARA |
granting_department |
Perpustakaan Tun Abdul Razak |
publishDate |
2013 |
url |
https://ir.uitm.edu.my/id/eprint/12024/1/TM_NASZARIAH%20MOHD%20NOOR%20EE%2013_5%201.pdf |
_version_ |
1783733161378709504 |