Comparing the effectiveness between Air-Q intubating laryngeal airway and ambu® auragain™ laryngeal mask for controlled ventilation in paediatric patients : a randomized controlled trial
Background: Supraglottic airway device (SAD) is a common device use in anaesthesia practice including paediatric patients. Air-Q ILA (Cook gas LLC; Mercury Medical, Clearwater, FL, USA) is the newer first generation of SAD that can use for both primary airway device and an aid for tracheal intuba...
Saved in:
Main Author: | |
---|---|
Format: | Thesis |
Language: | English |
Published: |
2018
|
Subjects: | |
Online Access: | http://eprints.usm.my/46743/1/Dr.%20Nirawanti%20Mohamad%20Said-24%20pages.pdf |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
id |
my-usm-ep.46743 |
---|---|
record_format |
uketd_dc |
spelling |
my-usm-ep.467432020-07-16T03:32:08Z Comparing the effectiveness between Air-Q intubating laryngeal airway and ambu® auragain™ laryngeal mask for controlled ventilation in paediatric patients : a randomized controlled trial 2018 Said, Nirawanti Mohamad RD Surgery Background: Supraglottic airway device (SAD) is a common device use in anaesthesia practice including paediatric patients. Air-Q ILA (Cook gas LLC; Mercury Medical, Clearwater, FL, USA) is the newer first generation of SAD that can use for both primary airway device and an aid for tracheal intubation. Available literature demonstrated that this device performed better and equally to the other SAD including second generation of SAD. Ambu® AuraGain™ (Ambu, Ballerup, Denmark) is a newer second generation of SAD which incorporates both integrating gastric port access and intubation capability. The study is conducted to compare the effectiveness between Air- Q and Ambu AuraGain for controlled ventilation in children up to 30kg. Methods: 64 paediatric patients underwent various short surgical procedures were randomly assigned to receive either an Air-Q or Ambu AuraGain. Fibreoptic (FO) grades of laryngeal view were measured as the primary outcome. The secondary outcomes measured were oropharyngeal leak pressure (OLP), number of attempts, time of successful insertion, quality of airway during placement and maintenance of anaesthesia, haemodynamic parameters and complications. Results: Air-Q has more favourable FO grades of view compared to the Ambu AuraGain (P = 0.047). OLP is significantly higher in Air-Q group compared to Ambu AuraGain (19.41 ± 1.19 cm H20 vs 17.56 ± 1.52 cm H20, P value = <0.001). There were no differences in term of number of attempts, time of successful insertion, quality of airway during placement and maintenance of anaesthesia and complications. 2018 Thesis http://eprints.usm.my/46743/ http://eprints.usm.my/46743/1/Dr.%20Nirawanti%20Mohamad%20Said-24%20pages.pdf application/pdf en public masters Universiti Sains Malaysia Pusat Pengajian Sains Perubatan |
institution |
Universiti Sains Malaysia |
collection |
USM Institutional Repository |
language |
English |
topic |
RD Surgery |
spellingShingle |
RD Surgery Said, Nirawanti Mohamad Comparing the effectiveness between Air-Q intubating laryngeal airway and ambu® auragain™ laryngeal mask for controlled ventilation in paediatric patients : a randomized controlled trial |
description |
Background: Supraglottic airway device (SAD) is a common device use in anaesthesia
practice including paediatric patients. Air-Q ILA (Cook gas LLC; Mercury Medical,
Clearwater, FL, USA) is the newer first generation of SAD that can use for both
primary airway device and an aid for tracheal intubation. Available literature
demonstrated that this device performed better and equally to the other SAD including
second generation of SAD. Ambu® AuraGain™ (Ambu, Ballerup, Denmark) is a newer
second generation of SAD which incorporates both integrating gastric port access and
intubation capability. The study is conducted to compare the effectiveness between Air-
Q and Ambu AuraGain for controlled ventilation in children up to 30kg.
Methods: 64 paediatric patients underwent various short surgical procedures were
randomly assigned to receive either an Air-Q or Ambu AuraGain. Fibreoptic (FO)
grades of laryngeal view were measured as the primary outcome. The secondary
outcomes measured were oropharyngeal leak pressure (OLP), number of attempts, time
of successful insertion, quality of airway during placement and maintenance of
anaesthesia, haemodynamic parameters and complications.
Results: Air-Q has more favourable FO grades of view compared to the Ambu
AuraGain (P = 0.047). OLP is significantly higher in Air-Q group compared to Ambu
AuraGain (19.41 ± 1.19 cm H20 vs 17.56 ± 1.52 cm H20, P value = <0.001). There were
no differences in term of number of attempts, time of successful insertion, quality of
airway during placement and maintenance of anaesthesia and complications. |
format |
Thesis |
qualification_level |
Master's degree |
author |
Said, Nirawanti Mohamad |
author_facet |
Said, Nirawanti Mohamad |
author_sort |
Said, Nirawanti Mohamad |
title |
Comparing the effectiveness between Air-Q intubating laryngeal airway and ambu® auragain™ laryngeal mask for controlled ventilation in paediatric patients : a randomized controlled trial |
title_short |
Comparing the effectiveness between Air-Q intubating laryngeal airway and ambu® auragain™ laryngeal mask for controlled ventilation in paediatric patients : a randomized controlled trial |
title_full |
Comparing the effectiveness between Air-Q intubating laryngeal airway and ambu® auragain™ laryngeal mask for controlled ventilation in paediatric patients : a randomized controlled trial |
title_fullStr |
Comparing the effectiveness between Air-Q intubating laryngeal airway and ambu® auragain™ laryngeal mask for controlled ventilation in paediatric patients : a randomized controlled trial |
title_full_unstemmed |
Comparing the effectiveness between Air-Q intubating laryngeal airway and ambu® auragain™ laryngeal mask for controlled ventilation in paediatric patients : a randomized controlled trial |
title_sort |
comparing the effectiveness between air-q intubating laryngeal airway and ambu® auragain™ laryngeal mask for controlled ventilation in paediatric patients : a randomized controlled trial |
granting_institution |
Universiti Sains Malaysia |
granting_department |
Pusat Pengajian Sains Perubatan |
publishDate |
2018 |
url |
http://eprints.usm.my/46743/1/Dr.%20Nirawanti%20Mohamad%20Said-24%20pages.pdf |
_version_ |
1747821721169166336 |