Photoacoustic imaging system for fluid flow visualization and biological tissues characterization
Photoacoustic (PA) imaging is a biomedical imaging modality, which work is based on the photoacoustic effect, to provide illumination of biological tissues with strong optical absorption contrast and high spatial resolution at a short scanning time. The drawbacks of some existing PA fluid imag...
Saved in:
主要作者: | |
---|---|
格式: | Thesis |
语言: | English English English |
出版: |
2020
|
主题: | |
在线阅读: | http://eprints.uthm.edu.my/1006/1/24p%20CHUA%20HUI%20LING.pdf http://eprints.uthm.edu.my/1006/2/CHUA%20HUI%20LING%20COPYRIGHT%20DECLARATION.pdf http://eprints.uthm.edu.my/1006/3/CHUA%20HUI%20LING%20WATERMARK.pdf |
标签: |
添加标签
没有标签, 成为第一个标记此记录!
|
id |
my-uthm-ep.1006 |
---|---|
record_format |
uketd_dc |
spelling |
my-uthm-ep.10062021-09-20T07:11:19Z Photoacoustic imaging system for fluid flow visualization and biological tissues characterization 2020-07 Chua, Hui Ling RC71-78.7 Examination. Diagnosis Including radiography Photoacoustic (PA) imaging is a biomedical imaging modality, which work is based on the photoacoustic effect, to provide illumination of biological tissues with strong optical absorption contrast and high spatial resolution at a short scanning time. The drawbacks of some existing PA fluid imaging systems, which include expensive equipment and their maintenance cost, limited sensitivity in detecting signals from restricted regions. This research describes an in-house developed two-axis PA imaging system for investigation of fluid flow and photoacoustic signatures of biological tissues using a continuous laser beam of output wavelength 633 nm to deliver light to the targeted tissue samples and fluid. The resulting acoustic signal detected by a transducer was used to determine the amplitude of tissues optical absorption via the measured phase value (Ф). This research began with the validation of the functionality of the developed system using biological system comprised of a mock circulatory system overlaid by different parts of poultry namely fat, liver and muscle of different sizes. The validated system was then deployed for use on human subjects, and the considered experiment settings included at rest, under warm water and arterial blood flow occlusion conditions. This work reported a consistent increase in the PA signals of all tissues with both sample size and the fluid flow rate. In addition, fat tissues were found to produce the largest PA signals with mean ±standard deviation (SD) Ф = 1.12 ± 0.11, while muscle produced the least signals with Ф = 0.828 ±0.20, which trend agreed well with the previous literature. It was found from the experiments on human subjects that phase difference (ΔФ) was proportional to the change in the velocity of blood flow within microcirculation of an investigated site. The mean and SD of percent relative phase difference for these volunteers were calculated as 51.68 % ± 24.27 % and -68.57 % ± 14.78 % for warm water and blood flow occlusion condition, respectively. The overall sensitivity of the system is 77. 3%. This work concluded the feasibility of this system for non-invasive assessment and visualization of blood perfusion and biological tissues, which deemed it suitable for implementation in healthcare applications. 2020-07 Thesis http://eprints.uthm.edu.my/1006/ http://eprints.uthm.edu.my/1006/1/24p%20CHUA%20HUI%20LING.pdf text en public http://eprints.uthm.edu.my/1006/2/CHUA%20HUI%20LING%20COPYRIGHT%20DECLARATION.pdf text en staffonly http://eprints.uthm.edu.my/1006/3/CHUA%20HUI%20LING%20WATERMARK.pdf text en validuser mphil masters Universiti Tun Hussein Onn Malaysia Fakulti Kejuruteraan Elektrik dan Elektronik |
institution |
Universiti Tun Hussein Onn Malaysia |
collection |
UTHM Institutional Repository |
language |
English English English |
topic |
RC71-78.7 Examination Diagnosis Including radiography |
spellingShingle |
RC71-78.7 Examination Diagnosis Including radiography Chua, Hui Ling Photoacoustic imaging system for fluid flow visualization and biological tissues characterization |
description |
Photoacoustic (PA) imaging is a biomedical imaging modality, which work is based
on the photoacoustic effect, to provide illumination of biological tissues with strong
optical absorption contrast and high spatial resolution at a short scanning time. The
drawbacks of some existing PA fluid imaging systems, which include expensive
equipment and their maintenance cost, limited sensitivity in detecting signals from
restricted regions. This research describes an in-house developed two-axis PA imaging
system for investigation of fluid flow and photoacoustic signatures of biological
tissues using a continuous laser beam of output wavelength 633 nm to deliver light to
the targeted tissue samples and fluid. The resulting acoustic signal detected by a
transducer was used to determine the amplitude of tissues optical absorption via the
measured phase value (Ф). This research began with the validation of the functionality
of the developed system using biological system comprised of a mock circulatory
system overlaid by different parts of poultry namely fat, liver and muscle of different
sizes. The validated system was then deployed for use on human subjects, and the
considered experiment settings included at rest, under warm water and arterial blood
flow occlusion conditions. This work reported a consistent increase in the PA signals
of all tissues with both sample size and the fluid flow rate. In addition, fat tissues were
found to produce the largest PA signals with mean ±standard deviation (SD) Ф = 1.12
± 0.11, while muscle produced the least signals with Ф = 0.828 ±0.20, which trend
agreed well with the previous literature. It was found from the experiments on human
subjects that phase difference (ΔФ) was proportional to the change in the velocity of
blood flow within microcirculation of an investigated site. The mean and SD of percent
relative phase difference for these volunteers were calculated as 51.68 % ± 24.27 %
and -68.57 % ± 14.78 % for warm water and blood flow occlusion condition,
respectively. The overall sensitivity of the system is 77. 3%. This work concluded the
feasibility of this system for non-invasive assessment and visualization of blood
perfusion and biological tissues, which deemed it suitable for implementation in
healthcare applications. |
format |
Thesis |
qualification_name |
Master of Philosophy (M.Phil.) |
qualification_level |
Master's degree |
author |
Chua, Hui Ling |
author_facet |
Chua, Hui Ling |
author_sort |
Chua, Hui Ling |
title |
Photoacoustic imaging system for fluid flow visualization and biological tissues characterization |
title_short |
Photoacoustic imaging system for fluid flow visualization and biological tissues characterization |
title_full |
Photoacoustic imaging system for fluid flow visualization and biological tissues characterization |
title_fullStr |
Photoacoustic imaging system for fluid flow visualization and biological tissues characterization |
title_full_unstemmed |
Photoacoustic imaging system for fluid flow visualization and biological tissues characterization |
title_sort |
photoacoustic imaging system for fluid flow visualization and biological tissues characterization |
granting_institution |
Universiti Tun Hussein Onn Malaysia |
granting_department |
Fakulti Kejuruteraan Elektrik dan Elektronik |
publishDate |
2020 |
url |
http://eprints.uthm.edu.my/1006/1/24p%20CHUA%20HUI%20LING.pdf http://eprints.uthm.edu.my/1006/2/CHUA%20HUI%20LING%20COPYRIGHT%20DECLARATION.pdf http://eprints.uthm.edu.my/1006/3/CHUA%20HUI%20LING%20WATERMARK.pdf |
_version_ |
1747830703000649728 |